27 Oct Tomographic ultrasound (tUS) for thyroid volume determination
Tomographic ultrasound (tUS) for thyroid volume determination
27 October 2026
Michael Pöhlmann, Martin Freesmeyer, Christian Kühnel, Falk Gühne, Philipp Seifert
Background
Accurate thyroid volume determination is essential for treatment planning.
This feasibility study evaluated the accuracy of a novel tomographic ultrasound (tUS) approach compared to 2D ultrasound, using contrast-enhanced computed tomography (ceCT) as reference standard.
Methods
In this prospective, single-center study, 50 patients undergoing neck ceCT were enrolled.
Thyroid volume was measured by 2D ultrasound (ellipsoid model), automatedvolume (tUS uncorrected), physician-corrected volume (tUS corrected) and ceCT.
Linear regression for proportional bias, log-transformed Bland–Altman analysis, clinical error metrics and correction duration were evaluated.

Results
2D ultrasound showed significant proportional bias (slope=-0.215, p<0.0001) and systematic underestimation (mean difference -6.0mL; geometric mean ratio 0.76).
Corrected tUS eliminated proportional bias (p=0.0557) and reduced bias (mean difference +0.8mL; geometric mean ratio 1.04).
Corrected tUS achieved lower MAE (4.42mL vs. 6.53mL) and MAPE (19.4% vs. 25.4%) than 2D ultrasound, placing 60.0% vs. 34.0% of cases within ±20% error limits.
In enlarged glands (n=19), corrected tUS maintained superior accuracy (MAE 5.82mL, MAPE 12.4%) versus 2D ultrasound (MAE 11.35mL, MAPE 26.5%).
Physician correction averaged 2:51minutes, independent of thyroid volume (r=-0.081, p=0.5784).
Conclusions
In this feasibility study, physician-corrected tUS reduced the size-dependent systematic underestimation observed with 2D ultrasound and yielded volumes closer to ceCT, particularly in enlarged glands, within a clinically feasible workflow.
Because the best-performing method required manual expert correction and the limits of agreement with ceCT exceeded the predefined clinical threshold, corrected tUS should be regarded as a promising semi-automated approach that requires further validation, including interobserver assessment, before it can be considered interchangeable with ceCT.