21 Jun Comparison of ultrasonographic versus infrared pupillary assessment
Abstract
Objectives
To evaluate the correlation between ultrasonographic and infrared pupillary assessments in critically ill patients, including neurocritically ill patients.
Design
Prospective, observational study.
Setting
Tertiary teaching hospital intensive care unit (ICU) in Montevideo, Uruguay.
Patients
Twenty-six adults patients with age 18 or older admitted to the intensive care unit with and without neurologic pathology. A total of 212 pupillary measures were made between ultrasonographic pupillary assessment (UPA) and infrared pupillary assessment (IPA).
Interventions
This was a study that utilized non-invasive (minimal risk) ultrasonographic and infrared pupillary assessment in patients admitted to the ICU. Time between UPA and IPA in a single patient was consistently less than 3 min.
Measurements and main results
There was a strong positive association between UPA and IPA (right eye [OD]: r = de 0.926, p-value < 0.001; left eye [OS], r = 0.965, p-value < 0.001), also observed in the group of neurocritically ill patients (OD: r = 0.935, p-value < 0.001; OS: r = de 0.965, p-value < 0.001). Taking IPA as reference measure, the percent error for all subjects was 2.77% and 2.15% for OD and OS, respectively, and for neurocritically ill patients it was 3.21% and 2.44% for OD and OS, respectively.
Conclusions
Ultrasonographic pupillary assessment is strongly correlated with infrared pupillary assessment in critically ill patients, including neurocritically ill patients. Ultrasonographic pupillary assessment is a quick, feasible, non-invasive method that allows accurate pupillary assessment, particularly neurologic function, in patients in whom a more precise measurement of the pupil is required or eye opening is not possible (e.g., periorbital edema due to traumatic brain injury).
Introduction
Pupillary assessment (i.e., bilateral evaluation of pupils size, shape, symmetry, and reactivity) is a cornerstone of the neurologic physical examination in the critically ill, particularly the neurocritically ill. Systematic pupillary assessments are routinely performed in the critically ill because they can render early signs of neurologic deterioration, which, in some cases, may be the only clinically obtainable sign.
The Brain Trauma Foundation’s guidelines for the management of severe traumatic brain injury acknowledge the evaluation of pupils’ size and reactivity to light as source for early prognostic signs of neurologic pathology [1].
There are reports of pupillary assessment as far back as 1929, when Otto Lowestein first developed a technique based on the analysis of simple, direct, visual evaluations [2]. Nonetheless, despite remarkable technologic advances and substantial improvements on the understanding of the central nervous system, the pupillary examination has not much changed during the last century.
Regarding the conventional, visual pupillary assessment, there is considerable intra- and inter-observer variability due to inconsistency on several factors, such as illumination of patient’s room, examiner’s visual acuity and experience, and intensity and technique of light stimuli. Therefore, alternative techniques have been proposed [3, 4].
The most accepted alternative technique is the infrared pupillary assessment (IPA), first described in 1958 by Lowestein himself, and then, as from 1993, extensively studied by Merlin Larson; by these means, both of Lowestein and Larson ended up elucidating the association between pupillary abnormalities and brainstem injuries or pharmacologic effects [5, 6].
In 2003, Taylor et al., using an infrared pupillometer, established an association between intracranial pressure and pupillary abnormalities in patients with acute brain injury [7]. In 2011, Chen et al. using the infrared pupillometer, described a significant inverse relationship between decreasing pupil reactivity and increasing intracranial pressure; the first evidence of pupil abnormalities occurred, on average, 15.9 h prior to the time of the peak of intracranial pressure [8].
Today, the IPA is being increasingly adopted as a routine part of the neurologic examination, supported by a growing body of literature demonstrating its reliability, accuracy, and ease of use. Automated pupillometry allows rapid, non-invasive, reliable, and quantifiable assessment of pupillary function which may allow rapid diagnosis of intracranial pathology that affects clinical decision-making [9, 10].
On the other hand, the first report of ocular ultrasound was made in 1956 [11]. Technologic advances in ultrasound devices have since allowed implementing ocular ultrasound in the evaluation of several ophthalmologic pathologies such as ocular trauma and intraocular foreign body identification [12,13,14]. However, the ultrasonographic evaluation of the pupillary diameter and pupillary light reflex has not been well studied, let alone implemented.
In cases when direct visualization of the pupil is not possible due to soft tissue injury that precludes eye opening, alternative techniques such as IPA and LED-based perimetric pupillometry have been proposed; unfortunately, most of these techniques do not overcome the physical barrier placed by the soft tissue, rely on advanced devices that are not generally available in emergency situations, and require of specialized technical support [15,16,17].
In this context, UPA is particularly useful because it offers a simple yet accurate alternative that can be performed with small, portable devices at the point of care.
To the authors’ knowledge, few reports have been published regarding UPA application and clinical relevance, mostly in emergency medicine and critical care medicine. Nevertheless, none of them have evaluated and compared UPA’s accuracy to standard techniques such as IPA [18,19,20,21].
Objective
To evaluate the correlation between ultrasonographic and infrared pupillary assessments in critically ill patients, including neurocritically ill patients.
Methods
Study design and study population
This was a prospective, observational study following the Helsinki Declaration and approved by the institutional review board. Patients older than 18 years who were admitted to the ICU of the Hospital de Clínicas, Montevideo, Uruguay, from October 1, 2018 to March 1, 2019 were included. Exclusion criteria were patients with ophthalmologic or periorbital pathology, including abnormal pupillary anatomy or neurologic assessment at baseline, or with facial or periorbital soft tissue edema that precluded eye opening, and therefore adequate IPA.
Upon admission to the ICU, consent was obtained from patients’ family members. Disease severity was determined following the Glasgow Coma Scale, and electronic medical records were reviewed to obtain demographics (age, gender, race), vital signs (heart rate, respiratory rate, blood pressure, temperature), and hemodynamic parameters (if any).
Non-invasive (minimal risk) ultrasonographic and infrared pupillary assessments were performed during same visit to the patient. Lights in patients’ room were turned off and windows and curtains were closed, thereby ensuring homogeny in rooms’ illumination regardless of time of the day.
Because UPA required manual measurement of pupillary diameter, we performed UPA first and then IPA, thereby reducing potential source of observer bias. The time frame between UPA and IPA was less than 3 min, during which there was no administration of new medications or physical maneuvers or activities on the patient. The examinations, both UPA and IPA, were performed at any time during ICU stay, by three authors (C.Y, S.P and M.L) who were experienced in UPA and IPA. On some patients, two or more pupillary examinations (UPA and IPA) were performed, each on different days while still in the ICU. The ultrasound and infrared devices and the information they yielded were used as standard of care; however, the devices themselves were not investigated.
Ultrasonographic pupillary assessment (UPA)
The ultrasonographic pupillary assessment included evaluation of bilateral pupillary diameter and pupillary light reflex. With the ultrasound machine on “small parts” preset and the patient in a supine, semi-recumbent position, a linear 7.5–15 MHz transducer was gently placed over the lower edge of the closed eye for a trans-palpebral tangential view, with the probe marker pointed toward the right side of the patient (Fig. 1).

Ultrasonographic pupillary assessment (UPA)
The infrared pupillary assessment included evaluation of bilateral pupillary diameter and pupillary light reflex. The infrared pupillometer is a monocular, stand-alone, hand-held, battery-operated instrument, which captures and analyzes the images in less than 3 s. When placed over the eye, the infrared pupillometer shows a light stimulus over the open eye in order to assess for the pupillary light reflex (Fig. 3).
The overall time for IPA per patient, including right and left eye, was never more than 1 min. The infrared pupillometer machine is designed to produce minimal risk to the user and the patient. The only designated mechanical contact point with the patient is the headrest. All levels of radiation fall below threshold values recommended by the International Commission on Non-Ionizing Radiation Protection.

For both UPA and IPA, normal values of pupillary diameter were 2 to 4 mm. A pupillary diameter less than 2 mm was documented as miosis, whereas a pupillary diameter greater than 4 mm was documented as mydriasis.
The pupillary light reflex was determined by the percentage of reduction in size of pupil diameter immediately after light stimuli. Such a light stimuli was given by the flashlight of a mobile phone model iPhone 6, iPhone 7, iPhone 8 or iPhone X, at an approximate distance of 8 to 12 cm from the eye, and at an angle of 90 degrees (i.e., perpendicular) to the coronal plane of the pupil, all of which have the same documented luminosity, regardless of remaining battery, according to Apple Inc. and were available to the research staff.
The reduction in pupillary diameter was calculated as follows: (maximum resting aperture − minimum aperture)/maximum resting aperture; where maximum resting aperture was the maximal pupillary diameter recorded before light stimuli was shown, minimum aperture was the minimal pupillary diameter recorded after light stimuli was shown. A normal pupillary light reflex was considered as percentage of reduction in size of pupil diameter after light stimuli of 10% or more according to a large study by Taylor et al. [7].
Statistical analysis
For the comparison between UPA and IPA, Bland–Altman and dispersion plots, simple linear regression models, and the Chi-square test were used. The Shapiro–Wilk test and Q–Q plots were used to corroborate normal distribution of the data. All calculations were done using the R programming language and statistical software (version 3.5.1).consolidation)